Alphasense A4
Represents calibration procedure parameters for SensorModel.ALPHASENSE_A4_CO, SensorModel.ALPHASENSE_A4_NO2, SensorModel.ALPHASENSE_A4_H2S gas sensors.
Inheritors
Types
Link copied to clipboard
This procedure measures the electronic offsets of the Working Electrode (WE) and Auxiliary Electrode (AE) in mV. Important: Before performing the calibration, ensure that the gas sensor is removed from the PCB.
Link copied to clipboard
data class SensorSensitivityCalibration(val gasConcentration: Double) : CalibrationProcedureParameters.AlphasenseA4
This procedure measures the sensor's sensitivity in nA/ppm. The procedure can be completed early using the SensorCommands.finishCalibration command. Important: All calibration parameters, except for sensor sensitivity, must be known before running the procedure: WE and AE electronic offsets, sensor offsets, and PCB gain.